Flow Regulator
· Specification and model
ConfigurationRef.NOOperating RangeRemark
Double Scale11502 20-250ml/hFollow the
insruction for use


· Product Features

1. Place the I.V.s solution container at about 80 cm above the outlet level;

2. Connect the I.V.set to the container;

3. Connect the Extension Set with Regulator to the I.V.set (not necessary if Regulator is already contained in I.V.set);

4. Open the clamp to begin priming the line;

5. Check that liquid is coming out of the end of the line;

6. Regulator is provided in the OPEN position;

7. PRIME Regulator completely turning Regulator from OPEN position to OFF position and then 

adjust the Regulator scale tothe required value;

8. Connect the I.V.line to the catheter or needle;

9. Double check that Regulator is delivering the required flow rate by counting the drops.